Compass zone adjustment

Compass calibration zones


Compass calibration zones

1. Refer to the “Compass calibration zones” map shown above to verify that the compass zone setting is correct for your geographical location.

2. Press and hold the switch for 6 seconds until the zone selection comes up (a number will be displayed in the mirror compass window).

3. Press the switch repeatedly until the correct zone setting for your location is displayed.

4. Releasing the button for 10 seconds will exit the zone setting mode.

    See also:

    Airflow control buttons
    Each airflow control button activates the following air outlets. : Instrument panel outlets : Instrument panel outlets and foot outlets : Foot outlets, both side outlets of the instrument p ...

    Ventilation
    Ventilation To force outside air through the instrument panel outlets: 1. Set the air inlet selection button to the OFF position. 2. Set the airflow control dial to the “” position. 3. ...

    When “PUSH EJECT” is displayed
    Press the eject button to unload the disc. Check the disc for damage or deformation, and also check that the correct disc is inserted. Do not try to unload the disc forcibly. Single CDs (i.e., 8 cm/3 ...